Temperature INOPs
INOP Message, Indication
T1
T2
T3
T4
,
,
,
INOPs
Tamb
INOPs
Tart
INOPs
Tcereb
INOPs
Tcore
INOPs
<Temp Label> Deactivated
INOP tone
<Temp Label> Equip Malf
Numeric is replaced by -?-
INOP tone
<Temp Label>NoTransducer
Numeric is replaced by -?-
INOP tone
<Temp Label> Overrange
Numeric is replaced by -?-
INOP tone
<Temp Label> Unplugged
INOP tone
Tesoph
INOPs
Tnaso
INOPs
Trect
INOPs
Tskin
INOPs
Ttymp
INOPs
Tven
INOPs
Tvesic
INOPs
SpO2 INOPs
INOP Message, Indication
<SpO₂ Label> Deactivated
INOP tone
<SpO₂ Label> Equip Malf
Numeric is replaced by -?-
INOP tone
What to do
<Temp Label>
See
INOPs
<Temp Label>
See
INOPs
<Temp Label>
See
INOPs
<Temp Label>
See
INOPs
<Temp Label>
See
INOPs
A Temp measurement label in the measurement device has been deactivated, either by connecting a
Pressure transducer in the shared Press/Temp socket, or by deactivating the label in the
Measurement Selection
window.
The measurement automatically disappears from the display.
To switch the measurement on again, either reconnect a Temp transducer or reactivate the
Measurement Selection
measurement label in the
Contact your service personnel.
The temperature hardware is faulty.
Make sure the Temp probe is connected to the MMS or module.
If you silence this INOP, the measurement will be switched off.
Try changing the application site of the transducer.
[The temperature is less than -1°C, or greater than 45°C.]
A Temp measurement label has been deactivated, either by unplugging a module, or by deactivating
Measurement Selection
the label in the
The measurement automatically disappears from the display.
To switch the measurement on again, either replug the module or reactivate the measurement label in
Measurement Selection
the
<Temp Label>
See
INOPs
<Temp Label>
See
INOPs
<Temp Label>
See
INOPs
<Temp Label>
See
INOPs
<Temp Label>
See
INOPs
<Temp Label>
See
INOPs
<Temp Label>
See
INOPs
What to do
The SpO
measurement label in the measurement device has been deactivated by deactivating the label
2
Measurement Selection
in the
To switch the measurement on again, reactivate the measurement label in the
Selection
window.
The MMS or module is faulty. Unplug and replug the MMS or module. If the INOP persists, contact
your service personnel.
window.
window.
window.
window. The measurement automatically disappears from the display.
4 Patient Alarms and INOPs
Measurement
75