4 Patient Alarms and INOPs
INOP Message, Indication
Some ECG AlarmsOff
Resp INOPs
INOP Message, Indication
Resp Equip Malf
Numeric is replaced by -?-
INOP tone
Resp Erratic
Numeric is replaced by -?-
Resp Leads Off
Numeric is replaced by -?-
INOP tone
NBP INOPs
INOP Message, Indication
!! Cuff Not Deflat
!!!Cuff Not Deflat
Numeric is displayed with a -?-
Severe yellow/red INOP tone
During this INOP, alarms cannot
be paused or switched off.
!! Cuff Overpress
!!!Cuff Overpress
Numeric is displayed with a -?-
Severe yellow/red INOP tone
During this INOP, alarms cannot
be paused or switched off.
NBP Deactivated
INOP tone
NBP Equip Malf
Numeric is replaced by -?-
INOP tone
NBP Interrupted
Numeric is replaced by -?-
INOP tone
NBP Measure Failed
Numeric may be replaced by -?-
INOP tone
74
What to do
This message appears (if configured to do so) when the on/off settings of the yellow arrhythmia
alarms differ from the current Profile.
What to do
Contact your service personnel. The RESP hardware is faulty.
The monitor has detected too many artifacts in the measured Resp signal. Check that the RA and LL
electrodes are correctly attached and have not dried out.
Not all the required leads for Resp monitoring are attached. Make sure that the RA, LA, and LL leads
are attached.
What to do
Remove the cuff from the patient. Make sure that the tubing is not kinked or twisted and that the
correct patient category is selected. Try repeating the measurement.
You can silence the INOP, but the INOP message remains visible until the next NBP measurement is
Stop All
started or the
SmartKey is selected.
[Adult or pediatric patients: The NBP cuff pressure has exceeded 15 mmHg (2 kPa) for more than
3 minutes.
Neonatal patients: The NBP cuff pressure has exceeded 5 mmHg (0.7 kPa) for more than 90 seconds.]
The NBP cuff pressure exceeds the overpressure safety limits. Remove the cuff from the patient. Make
sure that the tubing is not kinked or twisted and that the correct patient category is selected. Try
restarting the measurement.
You can silence this INOP, but the INOP message remains visible until the next measurement is
Stop All
started or the
SmartKey is selected.
The NBP measurement label in the measurement device has been deactivated by deactivating the label
Measurement Selection
in the
To switch the measurement on again, reactivate the measurement label in the
Selection
window.
Remove the cuff from the patient. The NBP hardware is faulty. Contact your service personnel.
You can silence this INOP, but the INOP message remains visible until the next measurement is
Stop All
started or the
SmartKey is selected.
Check the tubing and cuff for leakages or kinks. Check that you are using the correct cuff size and
placement, and that the correct patient category is selected. Try restarting the measurement.
If the INOP occurs repeatedly, contact your service personnel.
You can silence this INOP, but the INOP message remains visible until the next measurement is
Stop All
started or the
SmartKey is selected.
This INOP arises when the measurement needed longer than the maximum time for inflation,
deflation or the total measurement.
Check that you are using the correct cuff size and placement, and that the correct patient category is
selected. Try restarting the measurement.
You can silence this INOP, but the INOP message remains visible until the next measurement is
Stop All
started or the
SmartKey is selected.
Check the condition and suitability of the patient for NBP monitoring. Use another cuff to continue
measuring.
window. The measurement automatically disappears from the display.
Measurement