4 Patient Alarms and INOPs
INOP Message, Indication
No Central Monit.
INOP tone
No ECG at Central
OtherMon INOP
!! OtherMon INOP
!!! OtherMon INOP
Rem.AlarmDev.Malf.
INOP tone
Settings Malfunct
INOP tone
Speaker Malfunct
INOP tone
SRR Incompatible
INOP tone
SRR Interference
INOP tone
SRR Invalid Chan
INOP tone
SRR Malfunction
INOP tone
TAAP Disabled
INOP tone
Tele Config Unsupp
INOP tone
Tele Equip Malf
INOP tone
Tele Incompatible
Tele INOP
!! Tele INOP
!!! Tele INOP
Tele Unsupported
INOP tone
<Timer Label> Elapsed
INOP tone
T-Mon INOP
!! T-Mon INOP
!!! T-Mon INOP
INOP tone
TransportEquipment
Unsupported LAN
INOP tone
User I/F Malfunct
INOP tone
70
What to do
There is a problem with the communication to the network. Central monitoring is currently not
possible (no patient alarms or information). Check the connection. In case the connection is via a
telemetry device, the current telemetry use model does not support central monitoring. Contact your
service personnel.
The ECG measured with the monitor ECG is not being sent to the Information Center via the
telemetry device.
Another monitor assigned to the patient (e.g. a transport monitor) is in an INOP condition. Check the
detailed INOP information at the Information Center and resolve the INOP condition(s) at the other
monitor.
There is a problem with the connection to the remote alarm device. Contact your service personnel to
check the remote alarm device and its connections.
The monitor cannot use the predefined settings for monitoring. Contact your service personnel.
Contact your service personnel to check the speaker and the connection to the speaker.
The short range radio component has compatibility problems. Contact your service personnel.
The short range radio connection has interference from another device. Try using another channel.
The channel configuration of the Short Range Radio is invalid. Check channel and channel mask
configuration.
Malfunction in the short range radio device. If the INOP persists contact your service personnel.
The currently selected telemetry configuration on the monitor does not allow connection of telemetry
devices to the monitor.
Telemetry Workflow configuration not supported. Check revision and configuration of monitor and
central.
The telemetry device has a malfunction. Disconnect and reconnect the telemetry device. If the INOP
reappears, replace the telemetry device.
SRR-enabled telemetry device is not supported by this central software revision. Please check
configuration.
A telemetry device assigned to the patient is in an INOP condition. Check the detailed INOP
information at the Information Center and resolve the INOP condition(s) at the telemetry device.
This telemetry device is not supported for direct connection to the monitor.
The time has expired for the timer indicated in the INOP text. Clearing the timer clears the INOP.
A transport monitor assigned to the patient is in an INOP condition. Check the detailed INOP
information at the Information Center and resolve the INOP condition(s) at the transport monitor.
The equipment is ready for transport. See the
There is a problem with the communication to the network and central monitoring is currently not
possible. Check the connection. If the INOP persists, switch off the monitor and contact your service
personnel.
Perform a visual and functional check of all the monitor input devices. Contact your service personnel.
Equipment
window for details.