Download  Print this page

Advertisement

4 Patient Alarms and INOPs
Alarm Message
* Run PVCs High
** <SO₂ Label> High
** <SO₂ Label> Low
** <SpO₂ Label> High
** <SpO₂ Label> Low
**ST Multi <n>,<n>
**ST Multi
at Information Center
**STE <n>,<n>
** ST-<n> High
** ST-<n> Low
* SVT
*** Tachy (Pulse)
** Tblood High
** Tblood Low
** tcpO₂ High
** tcpCO₂ High
** tcpO₂ Low
** tcpCO₂ Low
* Tele Alarm
** Tele Alarm
*** Tele Alarm
66
From
Condition
ECG/Arrhythmia
A run of > 2 PVCs.
SvO
/SO
The the measured intravascular oxygen
2
2
saturation has exceeded the high limit.
SvO
/SO
The measured intravascular oxygen saturation
2
2
has fallen below the low limit.
SpO
The arterial oxygen saturation has exceeded the
2
high alarm limit.
SpO
The arterial oxygen saturation has fallen below
2
the low alarm limit.
ECG/ST
Two contiguous ST leads <n> and <n> have
exceeded elevation or depression limits for
more than 60 seconds. The limit violations for
both leads must be with respect to same limit;
either both above the high limit or both below
the low limit.
ECG/ST
The ST depression or elevation is outside of
the limit in two or more leads. Check on the
monitor
ECG/ST
Two contiguous leads <n> and <n> are above
their respective STE limits
ECG/ST
The ST elevation in lead <n> is higher than the
limit. Lead is not contiguous with any other
lead.
ECG/ST
The ST depression in lead <n> is lower than
the limit. Lead is not contiguous with any other
lead.
ECG/Arrhythmia
A run of supraventricular beats > the SVT run
limit and heart rate > the SVT HR limit.
SpO
, Press,
Pulse rate > the extreme tachy alarm limit.
2
cmResp
C.O.
The blood temperature value has exceeded the
high alarm limit.
C.O.
The blood temperature value has fallen below
the low alarm limit.
tcGas
The tcpO
high alarm limit.
tcGas
The tcpO
the low alarm limit.
an assigned
A telemetry device used for the patient is in an
telemetry device
alarm condition. Check the detailed alarm
information at the Information Center.
or tcpCO
value has exceeded the
2
2
or tcpCO
value has fallen below
2
2
Indication
numeric flashes, yellow alarm lamp,
short yellow audible alarm
numeric flashes and high alarm
limit is highlighted, yellow alarm
lamp, alarm tone
numeric flashes and low alarm limit
is highlighted, yellow alarm lamp,
alarm tone
numeric flashes and high limit is
highlighted, yellow alarm lamp,
alarm tone
numeric flashes and low limit is
highlighted, yellow alarm lamp,
alarm tone
numeric flashes, yellow alarm lamp,
alarm tone
(on monitor) numeric flashes,
yellow alarm lamp, alarm tone
numeric flashes, yellow alarm lamp,
alarm tone
numeric flashes and high alarm
limit is highlighted, yellow alarm
lamp, alarm tone
numeric flashes and low alarm limit
is highlighted, yellow alarm lamp,
alarm tone
numeric flashes, yellow alarm lamp,
alarm tone
numeric flashes, alarm limit is
highlighted, red alarm lamp, alarm
tone
numeric flashes, high alarm limit is
highlighted, yellow alarm lamp,
alarm tone
numeric flashes, low alarm limit is
highlighted, yellow alarm lamp,
alarm tone
numeric flashes, high alarm limit is
highlighted, yellow alarm lamp,
alarm tone
numeric flashes, low alarm limit is
highlighted, yellow alarm lamp,
alarm tone
yellow or red alarm lamp and alarm
tone

Advertisement

Table of Contents

Comments to this Manuals

Symbols: 0