4 Patient Alarms and INOPs
Alarm Message
* Irregular HR
** <iTemp Label> High
** <iTemp Label> Low
* Missed Beat
* More Alarms
** More Alarms
*** More Alarms
* Multiform PVCs
** NBP High
** NBP Low
* Non-Sustain VT
* OtherMon Alarm
** OtherMon Alarm
*** OtherMon Alarm
* Pacer Not Capt
* Pacer Not Pacing
* Pair PVCs
* Pause
*** <Press Label>
Disconnect
64
From
Condition
ECG/Arrhythmia
Consistently irregular rhythm (irregular R-R
intervals).
iTemp
The temperature value has exceeded the high
alarm limit.
iTemp
The temperature value has dropped below the
low alarm limit.
ECG/Arrhythmia
No beat detected for > 1.75*(the averaged R-R
interval), or no beat detected for > one second
if HR > 120 (For non-paced patients only).
other assigned
There is more than one physiological alarm of
devices
the corresponding severity active on the devices
that monitor this patient. For details, check the
information displayed at the Information
Center.
ECG/Arrhythmia
Two different shaped Vs detected, each
occurring at least twice within the last 300 beats
and at least once within the last 60 beats.
NBP
The measured NBP value is above the high
alarm limit.
s
d
m
,
, or
systolic, diastolic or mean pressure has crossed
the limit.
NBP
The measured NBP value is below the low
alarm limit.
s
d
m
,
, or
systolic, diastolic or mean pressure has crossed
the limit.
ECG/Arrhythmia
A run of PVCs < the V-Tach run limit and
ventricular heart rate > the V-Tach HR limit
another assigned
Another monitor used for the patient is in an
monitor
alarm condition. Check the detailed alarm
information at the Information Center.
ECG/Arrhythmia
No beat detected for > 1.75*(the averaged R-R
(paced patients
interval) with detected pace pulse(s). (For paced
only)
patients only).
ECG/Arrhythmia
No beat and pace pulse detected for >
(paced patients
1.75*(the averaged R-R interval). (For paced
only)
patients only).
ECG/Arrhythmia
Two ventricular beats between two non-
ventricular beats.
ECG/Arrhythmia
No beat detected for a period > the pause
threshold.
PRESS
The pressure is non-pulsatile and the mean
pressure is continuously less than 10 mmHg
(1.3 kPa). This alarm occurs only with arterial
pressures (P, ABP, ART, Ao, BAP, FAP, PAP,
UAP, P1, P2, P3, P4).
after the label indicates whether the
after the label indicates whether the
Indication
numeric flashes, yellow alarm lamp,
short yellow audible alarm
numeric flashes and high limit is
high-lighted, yellow alarm lamp,
alarm tone
numeric flashes and low limit is
high-lighted, yellow alarm lamp,
alarm tone
numeric flashes, yellow alarm lamp,
short yellow audible alarm
yellow or red alarm lamp and alarm
tone
numeric flashes, yellow alarm lamp,
short yellow audible alarm
numeric flashes and high limit is
highlighted, yellow alarm lamp,
alarm tone
numeric flashes and low limit is
highlighted, yellow alarm lamp,
alarm tone
numeric flashes, yellow alarm lamp,
short yellow audible alarm
yellow or red alarm lamp and alarm
tone
numeric flashes, yellow alarm lamp,
short yellow audible alarm
numeric flashes, yellow alarm lamp,
short yellow audible alarm
numeric flashes, yellow alarm lamp,
short yellow audible alarm
numeric flashes, yellow alarm lamp,
short yellow audible alarm
numeric flashes, red alarm lamp,
alarm tone