4 Patient Alarms and INOPs
Alarm Message
*** Brady (Pulse)
*** Bradypnea
** CCO High
** CCI High
** CCO Low
** CCI Low
** cmRR High
** cmRR Low
** CPP High
** CPP Low
*** Desat
*** Desat xx < yy
** EC10 Alarm
*** EC10 Alarm
at Information Center
* End AFIB
* End Irregular HR
** et<Agent Label> High
** et<Agent Label> Low
** etCO₂ High
** etCO₂ Low
** etO₂ High
62
From
Condition
SpO2, Press,
Pulse rate < the extreme brady alarm limit.
cmResp
cmRR
The respiration rate has dropped below the
bradypnea limit.
CCO
Continuous Cardiac Output or CC Index is
above the high alarm limit.
CCO
Continuous Cardiac Output or CC Index is
below the low alarm limit.
cmResp
The respiration rate has exceeded the high
alarm limit.
cmResp
The respiration rate has dropped below the low
alarm limit.
CPP
The CPP value has exceeded the high alarm
limit.
CPP
The CPP value has fallen below the low alarm
limit.
SpO
The SpO
2
desaturation alarm limit. xx denotes the lowest
measured value, and yy is the desaturation limit.
EC10 IntelliBridge A yellow (**) or red (***) patient alarm is
present on the IntelliBridge module. Check the
monitor display for more detailed alarm
information.
ECG/Arrhythmia
Atrial fibrillation no longer detected for the
configured Afib end delay time
ECG/Arrhythmia
Irregular HR no longer detected for the
configured irregular HR end delay time.
GM
The end tidal agent high alarm limit has been
exceeded.
GM
The end tidal agent value has fallen below the
low alarm limit
CO
, AGM, GM
The end tidal CO
2
exceeded.
CO
, AGM, GM
The end tidal CO
2
low alarm limit.
O
, AGM
The end tidal O
2
exceeded.
value has fallen below the
2
high alarm limit has been
2
value has fallen below the
2
high alarm limit has been
2
Indication
numeric flashes and alarm limit is
highlighted, red alarm lamp, alarm
tone
numeric flashes and alarm limit is
highlighted, red alarm lamp, alarm
tone
numeric flashes and high alarm
limit is highlighted, yellow alarm
lamp, alarm tone
numeric flashes and low alarm limit
is highlighted, yellow alarm lamp,
alarm tone
numeric flashes and high alarm
limit is highlighted, yellow alarm
lamp, alarm tone
numeric flashes and low alarm limit
is highlighted, yellow alarm lamp,
alarm tone
numeric flashes and high limit is
highlighted, yellow alarm lamp,
alarm tone
numeric flashes and low limit is
highlighted, yellow alarm lamp,
alarm tone
numeric flashes, red alarm lamp,
alarm tone
(on monitor) yellow or red alarm
lamp; the alarm text is defined by
the IntelliBridge device driver
yellow alarm lamp, short yellow
alarm tone
yellow alarm lamp, short yellow
alarm tone
numeric flashes and high limit is
highlighted, yellow alarm lamp,
alarm tone
numeric flashes and low limit is
highlighted, yellow alarm lamp,
alarm tone
numeric flashes and high limit is
highlighted, yellow alarm lamp,
alarm tone
numeric flashes and low limit is
highlighted, yellow alarm lamp,
alarm tone
numeric flashes and high limit is
highlighted, yellow alarm lamp,
alarm tone