Patient Alarms and INOPs

4
This chapter lists patient alarms alphabetically, and technical alarms (INOPs) arranged by the source of
the INOP and then alphabetically, irrespective of their priority. All alarms and INOPs are listed here;
the ones which can appear on your monitor will depend on the model and the individual options.

Patient Alarm Messages

The measurement labels and abbreviations for pressure, temperature, and SpO
in the individual chapters.
Some alarms may be shown at the Information Center in shortened form, when transferred through
IntelliVue Instrument Telemetry. These shortened alarm texts are included in the list and identified
with the note "at Information Center".
Alarm Message
* AFIB
** AGT Mix MAC > 3
*** Apnea
*** Apnea x:yy
*** Apnea >10 min
*** Asystole
** awRR High
** awRR Low
** <BIS Label> High
** <BIS Label> Low
From
Condition
ECG/Arrhythmia
Atrial fibrillation waveform detected
GM
An agent mixture has been detected and the
sum of the 2 agent MAC components + MAC
(N
O) is ≥ 3 MAC (uncorrected)
2
CO
, Resp,
Respiration has stopped for longer than the
2
Spirometry, GM
preset apnea time. "x:yy" denotes the Apnea
duration in minutes and seconds
ECG
No QRS detected for a period > the asystole
alarm threshold
CO
, Resp, AGM,
The airway respiration rate has exceeded the
2
GM
high alarm limit.
CO
, Resp, AGM,
The airway respiration rate has dropped below
2
GM
the low alarm limit.
BIS
The Bispectral Index value has exceeded the
high alarm limit.
BIS
The Bispectral Index value has dropped below
the low alarm limit.
alarms are explained
2
Indication
yellow alarm lamp, short yellow
alarm tone
yellow alarm lamp, alarm tone.
numeric flashes, red alarm lamp,
alarm tone
numeric flashes, red alarm lamp,
alarm tone
numeric flashes and high limit is
highlighted, yellow alarm lamp,
alarm tone
numeric flashes and low limit is
highlighted, yellow alarm lamp,
alarm tone
numeric flashes and high limit is
highlighted, yellow alarm lamp,
alarm tone
numeric flashes and low limit is
highlighted, yellow alarm lamp,
alarm tone