3 Alarms
Review Alarms Window
The
Review Alarms
information.
If configured to do so, each alarm is shown with the alarm limit active when the alarm was triggered
and the maximum value measured beyond this limit. The
monitor was switched on (after being switched off for longer than 1 minute) and any actions related to
switching alarms on and off, entering or leaving Standby mode, silencing alarms or changing the ECG
source.
When you select an item from the list, you can get additional information about that item. If you select
a high or low limit alarm in the list, the
you select an alert other than a high or low alarm, a help text window opens with more information.
Some items in the list are simply log items not related to a patient alert as such (for example, Alarms
On or Alarms Off). You cannot see any further information if you select one of these items. When you
close these windows you will return to the
The information in the
leave Demonstration Mode.
The
Review Alarms
is disabled, the pause pop-up keys are inactive.

Latching Alarms

The alarm latching setting for your monitor defines how the alarm indicators behave when you do not
acknowledge them. When alarms are set to non-latching, their indicators end when the alarm condition
ends. Switching alarm latching on means that visual and/or audible alarm indications are still displayed
or announced by the monitor after the alarm condition ends. The indication lasts until you
acknowledge the alarm.
Viewing the Alarm Latching Settings
To see the alarm latching setting for your monitor:
1
In the monitor's
Select
2
This setting can only be changed in Configuration Mode. You should be aware of the settings chosen
for your unit. There are three possible choices each for visual and audible latching, red only, red and
yellow, and off. These choices can be combined to give the following settings:
Visual Latching
Red&Yellow
Red&Yellow
Red&Yellow
Red Only
Red Only
Off
58
window contains a list of the most recent alarms and INOPs with date and time
Review Alarms
window pop-up keys appear when the window is opened. If alarm pause extension
Main Setup
menu, select
Alarm Settings
, and see the
Audible Latching
Red&Yellow
Red Only
Off
Red Only
Off
Off
Review Alarms
window will open to provide further data. If
Graphical Trends
Review Alarms
window.
window is deleted when a patient is discharged, and when you
Alarms
.
Visual Latching
and
Audible Latching
window also shows when the
settings.