Default Settings Appendix

25
This appendix documents the most important default settings of your monitor as it is delivered from
the factory. For a comprehensive list and explanation of default settings, see the Configuration Guide
supplied with your monitor. The monitor's default settings can be permanently changed in
Configuration Mode.
NOTE
If your monitor has been ordered pre-configured to your requirements, the settings at delivery will be
different from those listed here.

Alarm and Measurement Default Settings

The default wave speed for all measurements except Respiration is 25 mm/sec. For Respiration the
default wave speed is 6.25 mm/sec.
In the following tables, settings may be entered only once per table row if they are the same for all
patient categories.

Alarm Default Settings

Alarm Settings
Alarm Volume
Alarms Off
Pause Al. 5min
Pause Al. 10min
Auto Alarms Off
AlarmOffReminder
Visual Latching
Audible Latching
Alarm Reminder
Reminder Time
Alarm Sounds
RedAlarmInterval
Yel.Al. Interval
Factory Default H10/H20/H40
5
2 min
Enabled
Enabled
Off
Off
Red&Yellow
Red&Yellow
On
3 min
Traditional
10 sec
20 sec
H30/H31 (deviations from H10/H20/
H40)
Red Only
Off
25
317