24 Specifications
P
= partial pressure at ambient temperature and pressure, dry
ATPD
M3014A Mainstream CO2
M3014A Mainstream CO
CO2
Range
Accuracy
Resolution
Stability:
Short term drift
Long term drift
awRR
Range
Accuracy
Warm-up Time
Response Time
M3014A Sidestream CO2
M3014A Sidestream CO
CO2
Range
Accuracy
Resolution
Stability:
Short term drift
Long term drift
awRR
Range
Accuracy
Warm-up Time
Sample Flow Rate
Total System Response Time
Operating Temperature
310
Performance Specifications
2
0 to 150 mmHg (0 to 20.0 kPa)
after 2 minutes warmup:
For values between 0 and 40 mmHg: ±2.0 mmHg (±0.29 kPa)
For values from 41 to 70 mmHg: ±5% of reading
For values from 71 to 100 mmHg: ±8% of reading
The specifications are valid for standard gas mixtures, balance air, fully hydrated at 35°C, P
760 mmHg, flow rate = 2 l/min.
Numeric: 1.0 mmHg (0.1 kPa)
Wave: 0.1 mmHg (0.01 kPa)
±0.8 mmHg over four hours
Accuracy specification will be maintained over a 120 hour period
2 to 150 rpm
±1 rpm
2 minutes with CO
2
Less than 60 ms (with adult or infant reusable or disposable adapter)
Performance Specifications
2
0 to 150 mmHg (0 to 20.0 kPa)
after 2 minutes warmup:
For values between 0 and 40 mmHg: ±2.0 mmHg (±0.29 kPa)
For values from 41 to 70 mmHg: ±5% of reading
For values from 71 to 100 mmHg: ±8% of reading
For values from 101 to 150 mmHg: ±10% of reading
At respiration rates above 80 rpm, all ranges are ±12% of actual. The specifications are valid for gas
mixtures of CO
, balance N
2
Numeric: 1.0 mmHg (0.1 kPa)
Wave: 0.1 mmHg (0.01 kPa)
±0.8 mmHg over four hours
Accuracy specification will be maintained over a 120 hour period
2 to 150 rpm
±1 rpm
2 minutes with CO
2
50 ±10 ml/minute
3 seconds
0 to 40°C (32 to 100°F)
transducer attached for full accuracy specification
, dry gas at 760 mmHg within specified operating temperature range.
2
sensor attached for full accuracy specification
=
abs