CO2
The CO
2
Cor.1:2005 (except alarm system; alarm system complies with IEC 60601-2-49:2001).
M3015A Microstream CO2
M3015A Microstream CO
2
CO2
Range
Accuracy
Resolution
Stability
AwRR
Range
Accuracy
IPI
Range
Resolution
Warm-up Time
Rise Time
Sample Flow Rate
Gas Sampling Delay
Time
Sound Pressure
Total System Response
Time
Microstream CO2 Humidity Correction Factor
Either BTPS or ATPD can be selected as the humidity correction factor for the Microstream CO
readings. The formula for the correction calculation is:
where:
P
= partial pressure at body temperature and pressure, saturated
BTPS
measurement in the monitor, M3014A and M3015A complies with EN ISO 21647:2004 +
Performance Specifications
0 to 150 mmHg (0 to 20 kPa), or 20% CO
Up to 5 minutes during warmup: ±4 mmHg or 12%, whichever is greater
After 5 minutes warmup:
0 to 40 mmHg (0 to 5.3 kPa):±2.2 mmHg (±0.3 kPa)
Above 40 mmHg (5.3 kPa):±(5.5% + (0.08%/mmHg above 40 mmHg)) of
reading
These specifications are valid for 21% O
temperature, up to 60 rpm in adult mode and 100 rpm in neonatal mode.
Outside of these conditions the accuracy reaches at a minimum ±4 mmHg or
±12% of the reading, whichever is greater.
Numeric: 1.0 mmHg (0.1 kPa)
Wave: 0.1 mmHg (0.01 kPa)
Included in Accuracy specifications
0 to 150 rpm
0 to 40 rpm: ±1 rpm
41 to 70 rpm: ±2 rpm
71 to 100 rpm: ±3 rpm
>100 rpm: ±5% of reading
1 to 10
1
5 minutes for full accuracy specification
190 ms for neonatal mode
(measured with FilterLine H for neonatal)
240 ms for adult mode
(measured with FilterLine H for adult)
50 + 15/-7.5 ml/minute
Maximum:
3 seconds (2m sample lines)
6 seconds (4m sample lines)
Acoustic noise: <45 dBA
The total system response time is the sum of the delay time and the rise time.
24 Specifications
, whichever is lower
2
and N
balance, up to 35°C ambient
2
2
2
309