24 Specifications
Invasive Pressure Alarm Specifications
Invasive Pressure Alarm
Range
Specifications
Pressure
-40 to 360 mmHg
(-5.0 to 48 kPa)
Extreme High
Difference to high limit 0 to
25 mmHg
Clamping at -40 to 360 mmHg
Extreme Low
Difference to low limit 0 to
25 mmHg
Clamping at -40 to 360 mmHg
Pulse
25 to 300 bpm
Tachycardia
Difference to high limit 0 to
50 bpm
Clamping at 150 to 300 bpm
Bradycardia
Difference to low limit 0 to
50 bpm
Clamping at 25 to 100 bpm
Temp
Complies with EN 12470-4:2000. Specified without transducer.
Temp Performance Specifications
Temp
Range
Resolution
Accuracy
Average Time Constant
Temp Alarm Specifications
Temp High/Low Alarms
308
-1 to 45°C (30 to 113°F)
0.1°C (0.2°F)
±0.1°C (±0.2°F)
Less than 10 seconds
Range
-1 to 45°C (30 to 113°F)
Adjustment
-40 to 50 mmHg
2 mmHg (0.5 kPa)
>50 mmHg
5 mmHg (1 kPa)
5 mmHg steps (0.5 kPa)
5 mmHg steps (1.0 kPa)
5 mmHg steps (0.5 kPa)
5 mmHg steps (1.0 kPa)
Adult:
1 bpm steps (25 to 40 bpm)
5 bpm steps (40 to 300 bpm)
Pedi/Neo:
1 bpm steps (25 to 50 bpm)
5 bpm steps (50 to 300 bpm)
5 bpm steps
5 bpm steps
5 bpm steps
5 bpm steps
Adjustment
-1 to 30°C (30 to 86°F), 0.5°C (1.0°F) steps
30 to 45°C (86 to 113°F), 0.1°C (0.2°F) steps
Delay
max. 12 seconds
max. 14 seconds
max. 14 seconds