NBP Alarm Specifications
Systolic
Diastolic
Mean
NBP Overpressure Settings
Adult
Pediatric
Neonate
Invasive Pressure and Pulse
Complies with IEC 60601-2-34:2000/EN60601-2-34:2000.
Invasive Pressure Performance Specifications
Measurement Range
Pulse Rate
Range
Accuracy
Resolution
Input Sensitivity
Transducer
Frequency Response
Zero Adjustment
Range
Accuracy
Drift
Gain Accuracy
Accuracy
Drift
Non linearity and
Hysteresis
Overall Accuracy
(including transducer)
Volume displacement of CPJ840J6
Range
Adult: 30 to 270 mmHg (4 to 36 kPa)
Pedi: 30 to 180 mmHg (4 to 24 kPa)
Neo: 30 to 130 mmHg (4 to 17 kPa)
Adult: 10 to 245 mmHg (1.5 to 32 kPa)
Pedi: 10 to 150 mmHg (1.5 to 20 kPa)
Neo: 10 to 100 mmHg (1.5 to 13 kPa)
Adult: 20 to 255 mmHg (2.5 to 34 kPa)
Pedi: 20 to 160 mmHg (2.5 to 21 kPa)
Neo: 20 to 120 mmHg (2.5 to 16 kPa)
>300 mmHg (40 kPa) >2 sec
>300 mmHg (40 kPa) >2 sec
>150 mmHg (20 kPa) >2 sec
-40 to 360 mmHg
25 to 350 bpm
±1% Full Range
1 bpm
Sensitivity: 5 µV/V/mmHg (37.5 µV/V/kPa)
Adjustment range: ±10%
Load Impedance: 200 to 2000 Ω (resistive)
Output Impedance: ≤3000 Ω (resistive)
dc to 12 Hz or 40 Hz
±200 mmHg (±26 kPa)
±1 mmHg (±0.1 kPa)
Less than 0.1 mmHg/°C (0.013 kPa/°C)
±1%
Less than 0.05%/°C
Error of ≤0.4 %FS (@CAL 200 mmHg)
±4% of reading or ±4 mmHg (±0.5 kPa), whichever is greater
0.1 mm
Adjustment
10 to 30 mmHg: 2 mmHg (0.5 kPa)
>30 mmHg: 5 mmHg (1 kPa)
not user adjustable
3
/100 mmHg
24 Specifications