Respiration Alarm
Range
Specifications
High
Adult/pedi: 10 to 100 rpm
Neo: 30 to 150 rpm
Low
Adult/pedi: 0 to 95 rpm
Neo: 0 to 145 rpm
Apnea Alarm
10 to 40 seconds
Respiration from the CL Respiration Pod
For the general measurement specifications, refer to the IntelliVue Cableless Measurements
Instructions for Use.
The alarm specifications for alarms generated by the CL Respiration pod and indicated by the monitor
are listed in the table below.
cmResp Alarm Specifications
Respiration Rate
Bradypnea
Pulse
Tachycardia
Bradycardia
SpO2
Complies with ISO 9919:2005 / EN ISO 9919:2009 (except alarm system; alarm system complies with
IEC 60601-2-49:2001).
Measurement Validation
The SpO
2
measured with a CO-oximeter. Pulse oximeter measurements are statistically distributed, only about
two-thirds of the measurements can be expected to fall within the specified accuracy compared to CO-
oximeter measurements.
Range
5 to 60 rpm
Difference to high limit 0 to 20 rpm
Clamping at 5 to 50 rpm
30 to 220 bpm
Difference to high limit 0 to 50 bpm
Clamping at 150 to 220 bpm
Difference to low limit 0 to 50 bpm
Clamping at 30 to 100 bpm
accuracy has been validated in human studies against arterial blood sample reference
Adjustment
< 20 rpm: 1 rpm steps
≥ 20 rpm: 5 rpm steps
< 20 rpm: 1 rpm steps
≥ 20 rpm: 5 rpm steps
5 second steps
Adjustment
<20 rpm: 1 rpm steps
>20 rpm: 5 rpm steps
1 rpm steps
<20 rpm: 1 rpm steps
>20 rpm: 5 rpm steps
30 to 40 bpm: 1 bpm steps
40 to 220 bpm: 5 bpm steps
5 bpm steps
5 bpm steps
5 bpm steps
5 bpm steps
24 Specifications
Delay
max. 14 seconds
for limits from 0 to 20 rpm:
max. 4 seconds
for limits above 20 rpm: max.
14 seconds
301