ECG/Arrhythmia/ST Performance Specifications
Bandwidth
Bandwidth
when the ECG is transmitted from
a telemetry device via short range
radio
Differential Input Impedance
Common Mode Rejection Ratio
Electrode Offset Potential Tolerance
Auxiliary Current
(Leads off Detection)
Input Signal Range
ECG/Arrhythmia/ST/QT
Alarm Specifications
HR
Extreme Tachy
Extreme Brady
Run PVCs
PVCs Rate
Vent Tach HR
Vent Tach Run
Vent Rhythm Run
SVT HR
SVT Run
ST High
ST Low
STE Limit
QTc High
ΔQTc High
Diagnostic Mode
Extended Monitoring Mode
Monitoring Mode
Filter Mode
Diagnostic Mode
Extended Monitoring Mode
Monitoring Mode
Filter Mode
Range
15 to 300 bpm
maximum delay: 10 seconds
according to AAMI EC 13-1992
standard
Difference to high limit 0 to
50 bpm
Clamping at 150 to 300 bpm
Difference to low limit 0 to
50 bpm
Clamping at 15 to 100 bpm
None, fixed setting 2 PVCs
1 to 99 PVCs/minute
20 to 300 bpm
3 to 99 PVCs/minute
3 to 99 PVCs/minute
120 to 300 bpm
3 to 99 SV beats
-19.8 to +20 mm
-20 to +19.8 mm
-20 to +20 mm
200 ms to 800 ms
30 ms to 200 ms
Adult/neo/pedi: 0.05 to 150 Hz
Adult/neo/pedi: 0.5 to 150 Hz
Adult: 0.5 to 40 Hz
Neo/pedi: 0.5 to 55 Hz
Adult/neo/pedi: 0.5 to 20 Hz
Adult/neo/pedi: 0.05 to 40 Hz
Adult/neo/pedi: 0.5 to 40 Hz
Adult: 0.5 to 40 Hz
Neo/pedi: 0.5 to 40 Hz
Adult/neo/pedi: 0.5 to 20 Hz
>2 MΩ RA-LL leads (Resp)
>5 MΩ at all other leads (at 10 Hz including patient cable)
Diagnostic mode: >86 dB (with a 51 kΩ/47 nF imbalance).
Filter mode: >106 dB (with a 51 kΩ/47 nF imbalance).
±500 mV
Active electrode: <100 nA
Reference electrode: <900 nA
±5 mV
Adjustment
Adult:1 bpm steps (15 to 40 bpm)
5 bpm steps (40 to 300 bpm)
Pedi/Neo:1 bpm steps (15 to 50 bpm)
5 bpm steps (50 to 300 bpm)
5 bpm steps
5 bpm steps
5 bpm steps
5 bpm steps
Not adjustable by user
1 PVC
5 bpm
1 PVC
1 PVC
5 bpm
1 SV beat
0.2 mm
0.2 mm
0.2 mm
10 ms steps
10 ms steps
24 Specifications