24 Specifications

Measurement Specifications

See the "Default Settings Appendix" for a list of the settings the monitor is initially shipped with.
ECG/Arrhythmia/ST/QT
Complies with:
IEC 60601-2-25:1993 + A1:1999/EN60601-2-25:1995 + A1:1999
IEC 60601-2- 27:2005/EN60601-2-27:2006
IEC 60601-2-51:2003/EN 60601-2-51:2003
AAMI EC11/EC13:1991/2002
ECG/Arrhythmia/ST Performance Specifications
Cardiotach
PVC Rate
ST Numeric
QT Numeric
QTc Numeric
ΔQTc Numeric
QT-HR Numeric
Sinus and SV Rhythm Ranges
298
Range
Accuracy
Resolution
Sensitivity
Range
Resolution
Range
Accuracy
Resolution
Range
Accuracy
Resolution
Range
Resolution
Range
Resolution
Range - adult
Range - pediatric and neonatal
Brady
Normal
Tachy
Adult/pedi: 15 to 300 bpm
Neo range: 15 to 350 bpm
Note: for rates equal to or less than 15 bpm, the displayed heart
rate is 0.
±1% of range
1 bpm
≥200 µV
peak
0 to 300 bpm
1 bpm
-20 to +20 mm
±0.5 mm or 15%, whichever is greater
0.1 mm
200 to 800 ms
±30 ms
8 ms
200 to 800 ms
1 ms
-600 to +600 ms
1 ms
15 to 150 bpm
15 to 180 bpm
Adult: 15 to 59 bpm
Pedi: 15 to 79 bpm
Neo: 15 to 89 bpm
Adult: 60 to 100 bpm
Pedi: 80 to 160 bpm
Neo: 90 to 180 bpm
Adult: >100 bpm
Pedi: >160 bpm
Neo: >180 bpm