Download  Print this page

M4605a Battery Specifications - Philips MP2 Instructions For Use Manual

Intellivue patient monitor.
Hide thumbs

Advertisement

M4607A Battery Specifications
Temperature Range
Humidity Range
Battery Type
Safety
Electromagnetic Compatibility (EMC)
Communication Standard

M4605A Battery Specifications

The battery lifetime is 3 years from manufacturing date or 500 charge/discharge cycles.
M4605A Battery Specifications
Physical Specifications
W x D x H
Weight
Performance Specifications
Nominal Voltage
Rated Capacity at discharge C/5
Continuous Discharge Capability
Environmental Specifications
Temperature Range
Humidity Range
Battery Type
Safety
Electromagnetic Compatibility (EMC)
Communication Standard
Discharge 0 to 60°C (32 to 122°F)
Charge 0 to 60°C (32 to 122°F)
Storage and Transportation: -20 to 65°C (-4 to 140°F)
Operating: 15% to 95% Relative Humidity (RH)
Storage and Transportation: 5% to 95% Relative Humidity (RH)
Lithium Ion Manganese, 10.8 V, 1000 mAh
complies with UL 2054
complies with the requirements for FCC Type B computing Device, and EN
61000-4-2 and EN 61000-3-2
complies with the SMBus specification v1.1
149 x 89 x 19.8 mm (5.866 x 3.504 x 0.78 in)
490 g (1.08 lb) per battery
10.8 Volt
6000 mAh
6.5 A
Discharge 0 to 50°C (32 to 122°F)
Charge 0 to 50°C (32 to 122°F)
Storage and Transportation: -20 to 65°C (-4 to 140°F)
Operating: 15% to 95% Relative Humidity (RH)
Storage and Transportation: 5 % to 95 % Relative Humidity (RH)
Smart Battery 10.8 V, 6000 mAh, Lithium Ion
complies with UL 2054
complies with the requirements for FCC Type B computing Device, and
EN 61000-4-2 and EN 61000-3-2
complies with the SMBus specification v 1.1
24 Specifications