Download  Print this page

Environmental Specifications - Philips MP2 Instructions For Use Manual

Intellivue patient monitor.
Hide thumbs

Advertisement

24 Specifications

Environmental Specifications

The monitor may not meet the performance specifications given here if stored or used outside the
specified temperature and humidity ranges.
The monitor is protected against ingress of objects and fluids according to IEC 60529 IP32 (see
specifications below). Do not expose the monitor directly to heavy rain.
When the monitor and related products have differing environmental specifications, the effective
range for the combined products is that range which is common to the specifications for all products.
Item
Temperature Range
Temperature Range with
the Wireless IntelliVue
Adapter or IntelliVue
Instrument Telemetry
Wireless Network or when
charging the battery
Humidity Range
Altitude Range
Ingress Protection
1
Sufficient for flight altitudes up to 12,000 m with pressurized cabins
292
Condition
Operating
Storage including transportation
Operating
Operating
Storage including transportation
Operating
Storage including transportation
Monitor
External Power Supply (M8023A)
Battery Extension (865297)
Range
0 to 40°C (32 to 104°F)
-20 to 60°C (-4 to 140°F)
0 to 35°C (32 to 95°F)
15% to 95% Relative Humidity (RH)
5% to 95% Relative Humidity (RH)
-500 m to 3000 m (10000 ft)
1
-500 m to 4600 m (15000 ft)
IP32 (protected against ingress of water when the water is
dripping vertically and the monitor is tilted up to 15°and ingress
of foreign objects larger than 2.5 mm)
IP31(protected against the ingress of solid foreign objects 2.5 mm
in diameter or larger, and the ingress of water when the water is
dripping vertically) when rested on its rubber feet on a flat, level
surface.
IP32 when mounted as described in the Service Guide.
IP32 (protected against ingress of water when the water is
dripping vertically and the monitor is tilted up to 15°, not
protected against heavy rain.)

Advertisement

Table of Contents

Comments to this Manuals

Symbols: 0
Latest comments: