Download  Print this page

Advertisement

LED Indicators

The Battery Extension has two LED indicators. The power LED lights green when the Battery
Extension is connected to external power.
The battery charge led gives battery status information.
External power available
LED indication
green
yellow
red flashing (short on phase)
red flashing (long on phase)
off
External power not available
LED indication
yellow flashing (short on phase)
red flashing (short on-phase)
red flashing (long on-phase)
To remove the monitor with MMS Extension,
press the release lever and push the monitor
across to release the connection.
Status
Battery fully charged.
Battery charging.
Battery maintenance required.
Battery Extension malfunction.
No battery inserted in the Battery Extension.
Status
Battery Extension is charging the monitor battery (monitor is
switched off).
If the monitor is running, this indicates very low battery charge
(<10 minutes running time left). If the monitor is not running,
this indicates that battery maintenance is required.
Battery Extension cannot provide power to the monitor.
Either the Battery Extension needs charging, or it has a
malfunction.
1 Basic Operation
21

Advertisement

Table of Contents

Comments to this Manuals

Symbols: 0