13 Monitoring Invasive Pressure
1 Tubing to manometer
2 Syringe with heparinised solution
3 To pressure connector on monitor
4 Patient connection stoppered
5 Off
3
Move the syringe barrel in and raise the mercury to 200 mmHg (30 kPa). 200 mmHg is the
recommended calibration pressure.
In the
4
Select the calibration pressure from the list, for example 200 mmHg.
5
6
Select
When the monitor displays
7
manometer tubing, syringe and extra stopcock. We recommend you replace the transducer dome
and tubing with sterile ones.
Label the transducer with the calibration factor shown in the
8
setup menu.
Reconnect the patient and start measuring again.
9
Troubleshooting the Pressure Calibration
The status line lists the probable cause of an unsuccessful calibration.
Message
<Press Label> can't calibr.; equipmt
malf.
<Press Label> can't calibrate;out of
range
<Press Label> can't calibrate; no
transd.
198
menu, select
Setup <Press Label>
Confirm
to recalculate the calibration factor using the applied pressure.
<Press Label> mercury calibr. done at <Date and Time>
.
Cal. Press
Corrective Action
Contact your service department. The pressure hardware is
faulty.
Make sure that you have selected the value for
you are applying to the transducer, and repeat the calibration.
Make sure that the transducer is connected and try again.
, remove the
field in the pressure's
Cal. Factor
Cal. Press
that