Enabling Sequence Mode and Setting Up The
Sequence
In the
1
Select
2
Up to four measurement cycles can be set up which will run consecutively. For each cycle you can
set the number of measurements and the interval between them. If you want to run less than four
cycles in a sequence, you can set the number of measurements for one or more cycles to
3
Select each sequence in turn and select the number of measurements and the time interval between
the measurements.
To have measurements continue after the sequence, set the number of measurements for your last
4
cycle to
CAUTION
Be aware that, if none of the cycles are set to
measurement of the cycle.
When the NBP measurement mode is set to
changed.

Choosing the NBP Alarm Source

You can monitor for alarm conditions in systolic, diastolic and mean pressure, either singly or in
parallel. Only one alarm is given, with the priority of mean, systolic, diastolic.
In the
Setup NBP
Menu option
Sys.
Dia.
Mean
Sys & Dia
Dia & Mean
Sys & Mean
Sys&Dia&Mean
If
Mean
is not selected as alarm source (
a mean value, mean alarms will nevertheless be announced using the most recent mean alarm limits.
Check that the mean alarm limits are appropriate for the patient, even when not using mean as the
alarm source. When no value can be derived an
menu, select
Setup NBP
Mode
to open the
Setup Sequence
Continuous
and this cycle will run indefinitely.
menu, select
Al. from
Pressure value monitored
systolic
diastolic
mean
systolic and diastolic in parallel
diastolic and mean in parallel
systolic and mean in parallel
all three pressures in parallel
and select
from the pop-up menu.
Sequence
window.
Setup Sequence
, NBP monitoring will end after the last
Continuous
Sequence
, the repetition time for
and choose from:
Sys.
,
Dia.
, or
Sys & Dia
selected), but the monitor can only derive
NBP Measure Failed
11 Monitoring NBP
Auto
mode cannot be
INOP will be displayed.
Off
.
189