6 ECG, Arrhythmia, ST and QT Monitoring
Additional Message
Invalid Rhythm for QTc
High QT-HR
Small R Wave
Small T Wave
End of T not Detected
QT out of Range
QTc out of Range
QTc Erratic
Switching Individual QTc Alarms On and Off
Each QTc alarm can be switched off individually. To switch an alarm on or off, in the
Analysis
menu, select
Changing QTc Alarm Limits
Set the high alarm limits based on your assessment of the patient's clinical condition, unit protocols,
physician orders or medication specified limits.
1
In the
Select the appropriate setting.
2
Select
3
Select the appropriate setting.
4

Switching QT Monitoring On and Off

To switch all QT monitoring on or off, in the
between
On
156
Cause of Invalid QT Measurement
Not enough valid R-R intervals to generate QT-HR, the averaged
HR used for QTc calculation
QT-HR exceeds the specified upper limit of 150 bpm (for adults) or
180 bpm (for neonates and pediatrics)
R-wave of the signal is too small
T-wave of the signal is too small
End of the T-Wave cannot be accurately detected
QT measurement is outside the specified range of valid QT values
(200-800 msec)
QTc measurement is outside the specified range of valid QTc values
(200-800 msec)
QTc measurements are not stable
QTc Alarm
or
ΔQTc Alarm
Setup QT Analysis
menu, select
ΔHighLimit
.
and
Off
.
to toggle between
On
High Limit
.
menu, select
Setup QT Analysis
Setup QT
and
Off
.
to switch
QT Analys.