6 ECG, Arrhythmia, ST and QT Monitoring
R-wave peak at 0 msec
1
2
J point, for example, 48 msec
Difference = ST value
3
ST measurement point, for example, J + 60 msec
4
Isoelectric point set to -80 msec
5
P
6
7
Q
8
S
T
9
CAUTION
If using ST analysis, the ST measurement points need to be adjusted when you start monitoring, and if
the patient's heart rate or ECG morphology changes significantly, as this may affect the size of the QT
interval and thus the placement of the ST point. Artifactual ST segment depression or elevation may
occur if the isoelectric point or the ST point is incorrectly set.
Always ensure that ST measurement points are appropriate for your patient.
Setting ISO and J-Point Detection Mode
There are two modes for ISO and J-point detection:
measurement points yourself. In
To set the detection mode, in the
Select
1
Select
2
Adjusting ST Measurement Points
If the ISO and J-point detection mode is set to
Auto
mode, only the ST point can be changed.
144
Auto
Setup ST Analysis
ISO/J Pt
.
or
mode.
Auto
Manual
Auto
and
Manual
mode the measurement points will be determined automatically.
window,
, all three measurement points can be set. In
Manual
. In
Manual
mode you can set the