The ST point, J-point and ISO point can be hidden using the

Updating ST Baseline Snippets

ST analysis requires valid samples to measure and store a snippet. ST Snippets and ST values are
updated every minute. If there is artifact in the signal, it may take longer for an ST snippet and an ST
value to appear.
The first baseline is stored automatically after ST monitoring is started, or when a new patient is
admitted.
To update ST baselines,
1
Select the ST numerics then
In the
2
all previously-stored baselines.

About the ST Measurement Points

The ST value for each beat complex is the vertical difference between the ISO point and the ST point,
as shown in the diagram below. The isoelectric (ISO) point provides the baseline, the ST point is at the
midpoint of the ST segment. The J point is where the QRS complex changes its slope; as it is a fixed
distance away from the ST point, it can be useful to help you position the ST point correctly.
window opens with the pop-up keys
/
Hide Points
,
Record STE
ST View
ST View
window, select
Update Baseline
6 ECG, Arrhythmia, ST and QT Monitoring
/
Current
Baseline
,
ST View
,
ST Map
,
Setup ST
Hide Points
to open the
ST View
window.
to store all current snippets as baselines. This deletes
/
(to select viewing
Overlap
and arrow keys for scrolling
pop-up key.