Intermittent Bundle Branch Block
Bundle branch and the other fascicular blocks create a challenge for the arrhythmia algorithm. If the
QRS during the block changes considerably from the learned normal, the blocked beat may be
incorrectly classified as ventricular, causing false PVC alarms. You should always select a lead where
the bundle branch block beats have an R-wave that is as narrow as possible to minimize incorrect calls.
Ventricular beats should look different from these 'normal beats'. Instead of trying to select two leads
with a narrow R-wave, it may be easier to just select one lead and use single lead arrhythmia
monitoring. Extra vigilance is required by the clinician for this type of patient.

Understanding the Arrhythmia Display

Your monitor screen may look slightly different from the illustration.
Beat label
1
Pace pulse marks
2
Rhythm status message
3
PVC Numeric
4
5
HR Numeric
Ectopic status message
6
Delayed arrhythmia wave
7
Viewing Arrhythmia Waves
To review arrhythmia beat labels, in the
The wave showing the primary ECG lead will be delayed by six seconds and shown on a gray
background. Beat labels will be annotated above the ECG wave and
To return to the normal ECG primary lead display, select
Arrhythmia Beat Labels
Arrhythmia beat labels tell you how the monitor is classifying beats.
Beat Label
Arrhythmia On
N
V
S
P
6 ECG, Arrhythmia, ST and QT Monitoring
menu, select
Setup Arrhy
Beat Classification
Normal
Ventricular Ectopic
Supra-ventricular Premature
Paced
.
Annotate Arrhy
Delayed
will be written beside it.
again.
Annotate Arrhy
Beat Label
Cardiotach Mode
B
B
B
B
131