Download  Print this page

Using Ecg Alarms - Philips MP2 Instructions For Use Manual

Intellivue patient monitor.
Hide thumbs

Advertisement

Alarm
** Pacer Not Capt
**Pacer Not Pacing
** SVT
* HR High
* HR Low
* AFIB
* End AFIB
** Irregular HR
* End Irregular HR
Note:
** HR Low
HR Low
and

Using ECG Alarms

ECG alarms can be switched on and off and the high and low alarm limits changed just like other
measurement alarms, as described in the "Alarms" chapter. Special alarm features which apply only to
ECG are described here.
Extreme Alarm Limits for Heart Rate
The extreme rate alarms, extreme tachycardia and extreme bradycardia, are generated by the active
alarm source, either HR or Pulse.
The difference between the low HR alarm limit and the extreme bradycardia limit is unit configured.
For example, if the low alarm limit is 60 bpm and the extreme bradycardia limit difference is
configured to be 20 bpm, then the extreme bradycardia limit is 40 bpm. If the difference is configured
to be 0, there will always be an extreme bradycardia alarm when the HR falls below the HR low limit.
Description
No beat detected for a period > 1.75
times the average R-R interval and pace
pulse(s) detected. (Paced mode on)
No beat detected for a period > 1.75
times the average R-R interval and no
pace pulse(s) detected (Paced mode on)
A run of consecutive beats labeled as S
with run length ≥ SVT run limit and
ventricular HR > SVT HR limit
Heart rate > the high HR limit
Heart rate < the low HR limit
An irregular rhythm of beats labeled as
N and variability in P-R intervals and
P-wave variability (For adult patient
category only)
Atrial Fibrillation no longer detected
for the Afib end delay time (For adult
patient category only)
An irregular rhythm of beats labeled as
N (irregular R-R intervals changes
greater than 12.5%)
Irregular HR rhythm no longer
detected for the irregular HR end delay
time
and
** HR High
are available only if HR alarms are configured as yellow alarms, and *
* HR High
are available only if HR alarms are configured as short yellow alarms.
6 ECG, Arrhythmia, ST and QT Monitoring
Arrhythmia Configuration
Arrhythmia
Basic
Off
Arrhythmia
x
x
x
(see note)
x
(see note)
Enhanced
Arrhythmia
x
x
x
x
(see note)
x
(see note)
x
x
x
x
*
125