5 Managing Patients and Equipment
ST
QT
SpO
T
2
1
ECG On/Off setting is forced to ON by the monitor, if the local setting is ON.
2
With PIIC, QT baseline is synchronized, but not QT snippets.
With the IntelliVue Information Center iX, the following additional measurement settings can be
PIIC iX
synchronized:
ECG/Arrhythmia
ST
NBP
Resp
SpO
2
1
Measurement Mode and Repetition Time can only be synchronized if SpO
telemetry device.
WARNING
Not all settings are synchronized; after changing the measurement source, always check that the
settings are appropriate.
If later the patient is disconnected from the other device, and reconnected to the monitor again, any
changes in the settings made in the meantime will be passed on to the monitor. In this way, settings
continuity is preserved when the measurement source changes.
NOTE
Settings synchronization can be switched off at the monitor in Configuration mode.
In certain situations, you will be asked to confirm that synchronization of settings is appropriate.
Settings synchronization can only take place when there is no patient information mismatch between
the monitor and the Information Center.
If a
Check ECG Settings
PIIC
the paced setting, are appropriate for your patient.
If a
Check ECG Settings
PIIC iX
the settings, especially the paced setting for ECG, are appropriate for your patient.
102
ST Analysis On/Off, ST Alarm On/Off, ISO Point, J point, ST point, ST
Alarm Limits, ST-Index On/Off, ISO/J-Point Detection, ST Lead On/
Off, ST Baseline
QT analysis On/Off, QT Lead, QTc High Alarm On/Off, ΔQTc Alarm
On/Off, QTc High Limit, ΔQTc High Limit, QT Baseline
SpO
Alarms on/off, SpO
2
High Alarm Limit, Desat Alarm Limit
NBP Alarm Suppression On/Off, Pulse(SpO
Some ECG Alarms INOP On/Off, Hexad Reconstruction Mode
STE On/Off, STE Alarm On/Off
Sys/Dia/Mean Alarm Limits, Alarms On/ Off, Alarm Source
Apnea Time, Alarm Limits, Alarm On/Off, Resp On/Off
SpO
Alarms on/off, SpO
2
Suppression On/Off, Pulse(SpO
1
Repetition Time
or
!! Check Pairing
INOP appears always check that the ECG settings, especially
,
Chk SpO₂T Settings
, or another check settings INOP appears, always check that
Alarm limits, SpO
Low Alarm Delay, SpO
2
2
) On/Off
2
Alarm limits, Desat Alarm Limit, NBP Alarm
2
) On/Off, Measurement Mode
2
comes from a Cableless SpO
2
2
2
1
,
Pod or a
2