Set Filter Status Procedure

 Press "Support" and log on using IST
 Click: Test/Utilities  Service Messaging  Check Filter Status
 Enter the
"Next Filter Cleaning
to enter the interval length.
 Clicking "Done" sets the date.
 You should verify that the
updated appropriately.

System Lift Tool and Servicing Casters

The recommended interval for caster maintenance is also every 12
months. You will need to use the system cart lift tool and Molygraph
SL3144 bearing grease as recommended by Philips if maintence or
replacement is needed.
Move the system, check caster operation and visually inspect for
damage.
Only the front casters have brake/steering locks. The back casters
swivel for steering. The single pedal at the front of the system
controls the brake and steering. The pedal has three positions:
 Fully up – The casters are steering locked; the system rolls freely
but does not steer.