To Test Foot Switch Operation; System Test; To Perform The System Test - Philips epiq 5 Training Manual

Ultrasound.
Hide thumbs

Advertisement

The System Test should be
run any time you suspect
problems with the system
or when a system error is
displayed.
Running the System Test
requires you to reboot the
system.
focus and that the gray bar displays the full range of gray
shades as seen on the monitor.
8. Verify that the print image is spatially correct, horizontally and
vertically.

To Test Foot Switch Operation:

1. Verify the foot switch is connected to the system.
2. Press each of the three pedals, and verify that they operate as
intended.

System Test

The System Test is a comprehensive test of the system operational
status and normally requires about 30 minutes.
This test has numerous subtests:
 System Functional Test
 All of the Field- Replaceable Unit (FRU) Comprehensive Tests
 All of the Interface Tests
If a subtest should fail, the system still completes all the remaining
subtests. System Test is available to all users and displays a pass-fail
result on the system monitor. The subtest results are annotated in
the system logs.
If an error message is displayed during the test, you will need to
reboot the system using the On/Off switch.
After you have run the system test, you should export the system test
results and other logs. You can use the Export Logs capability to
transfer encrypted error logs to either a DVD or a USB device. If the
data is not saved, an alert is displayed, advising you to check the
media in the drive and try again.

To perform the system test:

 Disconnect all transducers.
Press Support
© 2016 Conquest Imaging
EPIQ 5 & 7 Training Manual
Test/Utilities
tab.
61