4. Click Done.
5. Continue exercising the system as described in the other
performance-test procedures in this section.
6. When the test is finished, do the following:
a. Touch
End
Exam.
b. Edit the information in the
c. Click
End
Exam.
d. Click Done.

To Start an Exam when Data Security is Disabled:

1. Connect and initialize any transducer. If possible, connect
transducers to all four ports.
2. Touch
Patient
exam was ended.
3. Enter the patient data and click Done.
4. Select a transducer from the touch screen.
5. Touch an application/preset.
Now you can continue to exercise the system as described in the
following performance-test procedures in this section.

2D Imaging Test

Successful execution of this test verifies that the
controls function correctly in 2D imaging mode.

Test 2D operation:

1. Touch
2D
to enter 2D imaging (if needed.)
2. Turn 2D to adjust the gain.
3. Adjust the TGC slide controls. Verify the changes in overall
image brightness.
4. Turn Zoom. Verify that the image increases from and
© 2016 Conquest Imaging
EPIQ 5 & 7 Training Manual
Patient Data
form.
and click
End Exam
to ensure the previous
control panel
51