Download  Print this page

Power Distribution Board; Power Distribution Board Connectors; Temperature Control - Philips epiq 5 Training Manual

Ultrasound.
Hide thumbs

Advertisement

Power Distribution Board

The Power Distribution Board contains no active circuitry. It receives
DC voltages from the VPPM and supplies to system via the following
connectors that are listed as they appear from right to left when
viewed from the back of the system:
 Solenoid Control Module power connector
 Three standard PC ATX connections
 Backplane power connector
 Backplane signal connector

Power Distribution Board Connectors

Temperature Control

System temperature is monitored and controlled by the I
several Programmable Intelligent Processor (PIC) chips located in key
areas of the system. It controls air flow by altering the fan speeds as
needed. If any monitored location in the system still exceeds the
specified temperature tolerances after the fan speed is increased the
system will be shut down in order to avoid damage.
© 2016 Conquest Imaging
EPIQ 5 & 7 Training Manual
2
C bus and
29