Installation And Mounting; Jumper Rs 485 - Emerson CSI 6500 Quick Start Manual

Machinery health monitor a6151, dual channel low frequency shaft vibration monitor
Hide thumbs Also See for CSI 6500:
Table of Contents

Advertisement

Available languages

Available languages

Product Information / Produktinformation A6151
Shaft Vibration Monitor
1.5

Installation and Mounting

A prepared 3U slot in a 19" rack, or other Intermas−compatible enclosure, is required for
mounting the A6151 monitor. The slot must be fitted with a 48−pin plug connector (DIN 41612,
design F 48 M). The pin assignment is listed in the following table.
2
UN+ (+ 24V)
4
A (RS 485)
6
Supply2+ (sensor)
8
AIN2− (Input)
10
AIN2+ (Input)
12
NGL2 (scaled dc−output)
14
EO1 (voltage output1)
16
EO2 (voltage output2)
18
GWM (limit value)
20
NC
22
KEY−N (key signal input)
24
SC−A (oper. principle alert)
26
D1−C (danger1 collector)
28
D1−E (danger1 emitter)
30
D2−C (danger2 collector)
32
D2−E (danger2 emitter)

1.5.1 Jumper RS 485

RS 485 bus operation requires an electrical terminator on the first and last bus device.
This is done with plug−in jumpers, that are on the controller board. The figure shows the position
of the jumpers.
To activate the bus terminator and to place lines "A" and "B" on the references, plug the jumpers
as shown in Fig. b). Fig. a) shows the jumper position for deactivated bus termination and open
references (delivery status).
Plug−in jumper 1−2 closed:
Plug−in jumper 3−4 closed:
Plug−in jumper 5−6 closed:
d
U− (0V/Common)
GND (BP, Common, RS 485) B (RS 485)
Supply1− (sensor)
Supply2− (sensor)
GND
GND
EI1 (voltage input1)
EI2 (voltage input2)
I1− (current output common)
I2− (current output common)
GND
GND
A1−C (alert1 collector)
A1−E (alert1 emitter)
A2−C (alert2 collector)
A2−E (alert2 emitter)
Bus line "B" via pull−down resistor on ground
120 Ω resistor between "A" and "B"
Bus line "A" via pull−up resistor on +5 V
b
UB+ (+24V, redundant)
Supply1+ (sensor)
AIN1− (Input)
AIN1+ (Input)
NGL1 (scaled dc−output)
AC1 (sensor raw signal1)
AC2 (sensor raw signal2)
I1+ (current output CH1)
I2+ (current output CH2)
ES (external lock)
SC−D (oper. principle danger)
CC1−C (ch. clear1 collector)
CC1−E (ch. clear1 emitter)
CC2−C (ch. clear2 collector)
CC2−E (ch. clear2 emitter)
Page 10
z
2
4
6
8
10
12
14
16
18
20
22
24
26
28
30
32

Advertisement

Table of Contents
loading

Table of Contents