6 BASICS OF THE MEASUREMENT ................7 6.1 P G 1700 ..............7 ROBE ACCURACY DEVICE ACCURACY 6.2 P G 1700 / -20 / -30 ..........8 ROBE ACCURACY DEVICE ACCURACY 6.3 P ......................8 OSSIBLE ERRORS 7 CONFIGURATION......................8 8 ADJUSTMENT OF TEMPERATURE INPUT .............
H86.0.0X.6C-02 Operating Manual G 1700 Series page 3 of 14 _____________________________________________________ _____________________________________________________________________________ 2 Safety 2.1 Intended Use The device is covered to measure temperature in different mediums. G 1700: To connect different Pt1000 probes, the device is equipped with a BNC socket. By selecting a suitable temperature probe, it is possible to use the device in different applications.
H86.0.0X.6C-02 Operating Manual G 1700 Series page 4 of 14 _____________________________________________________ _____________________________________________________________________________ 2.2 Safety signs and symbols Warnings are labeled in this document with the followings signs: Caution! This symbol warns of imminent danger, death, serious injuries and significant damage to property at non-observance.
– here the highest system accuracy can be achieved. To optimize the accuracy, an offset and a slope correction can be executed at the G 1700 (take a look at chapter 8 Adjustment of temperature input). By the devices with fixed connected temperature probes, it is not necessary.
H86.0.0X.6C-02 Operating Manual G 1700 Series page 6 of 14 _____________________________________________________ _____________________________________________________________________________ 4 Operation 4.1 Display elements 1 Battery symbol: Rating of battery state Units display: Measured value units or display for “min/max/hold” Main display: Current temperature measured value or value ”min/max/hold”...
After that the device is ready for measuring. 6 Basics of the measurement 6.1 Probe accuracy/device accuracy G 1700 The G 1700 can be equipped with different interchangeable probes. The temperature probes are classified into the following classes: Class Maximum deviation Area of validity ±...
To receive high exchange accuracy without necessity of additional adjustment we recommend the usage of temperature probes from the class “A” or “AA”. 6.2 Probe accuracy/device accuracy G 1700 / -20 / -30 The display unit and the probe of instruments with fixed mounted probes have been matched very accurate ex works, see chapter 13 Specification.
Page 9
Min-alarm limit AL.LO Alarm starts when measuring value falls below the limit AL.K, (* or -94.0 °F - on G 1700 from -200.0 °C or -328.0 °F AL.LO … Max-alarm limit AL.K, Alarm starts when measuring value exceeds the limit 250.0 (*...
H86.0.0X.6C-02 Operating Manual G 1700 Series page 10 of 14 _____________________________________________________ _____________________________________________________________________________ 8 Adjustment of temperature input Invoke menu: device switched hold release Hold the key “down“during switching on the device until “T.OF“is shown in the auxiliary display. The menu can be used like described in chapter 7 Configuration.
H86.0.0X.6C-02 Operating Manual G 1700 Series page 11 of 14 _____________________________________________________ _____________________________________________________________________________ 9 Accuracy check / adjustment service You can send the device to the manufacturer for adjustment and inspection. Calibration certificate - DKD certificate - official certifications: If the measuring instrument is supposed to receive a calibration certificate, it has to be sent to the manufacturer (declare test points, e.
H86.0.0X.6C-02 Operating Manual G 1700 Series page 12 of 14 _____________________________________________________ _____________________________________________________________________________ 11 Fault and System Messages Error messages for measurement Description What to do? No display or Battery empty Replace battery (see chapter 10 confused Replacing batteries) characters, System error...
H86.0.0X.6C-02 Operating Manual G 1700 Series page 13 of 14 _____________________________________________________ _____________________________________________________________________________ 12.2 Disposal Dispense exhausted batteries at destined gathering places The device must not be disposed in the unsorted municipal waste! Send the device directly to us (sufficiently stamped), considering the above if it should be disposed.
Page 14
H86.0.0X.6C-02 Operating Manual G 1700 Series page 14 of 14 _____________________________________________________ _____________________________________________________________________________ Directives and standards The instruments confirm to following European directives: 2014/30/EU EMV directives 2011/65/EU RoHS Applied harmonized standards: EN 61326-1:2013 emissions level: class B emi immunity according to table 2 additional error: <...