Installation; Installation And Alignment - Miele PG 8583 Operating Instructions Manual

Laboratory glassware washer for laboratory glassware and utensils
Table of Contents

Advertisement

Installation

146

Installation and alignment

Please refer to the installation diagram provided.
 In order to reduce the risk of water damage, the area around the
machine should be limited to furniture and fittings that are designed
for use in commercial environments.
The machine must be stable and horizontal.
You can compensate for any unevenness in the floor level and height
of the machine by adjusting the four feet. The feet can be screwed
out to a maximum of 2 3/8" (60 mm).
 Do not lift the machine by protruding parts such as the control
panel.
They could be damaged or torn off.
 Some metal parts pose a risk of injury/being cut.
Wear cut-resistant protective gloves when transporting and setting
up the machine.
 For transport by means of a hand truck, the machine must be in
its original packaging or placed on a stable, continuous support.
Otherwise, components in the base of the machine can be
damaged.
The machine is suitable for the following types of installation:
– Freestanding.
– Slot-in:
The machine can be installed beside other machines or furniture or
in a suitable niche. The niche must be at least 23 5/8" (600 mm)
wide and 23 5/8" (600 mm) deep.
– Building under:
The machine can be built under a continuous worktop or the
draining board of a sink. The space provided must be at least
23 5/8" (600 mm) wide, 23 5/8" (600 mm) deep and
32 5/16" (820 mm) high.