Download  Print this page

Philips MX400 Instructions For Use Manual Page 96

Intellivue series patient monitors.
Hide thumbs
   
1
2
Table of Contents
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
27
28
29
30
31
32
33
34
35
36
37
38
39
40
41
42
43
44
45
46
47
48
49
50
51
52
53
54
55
56
57
58
59
60
61
62
63
64
65
66
67
68
69
70
71
72
73
74
75
76
77
78
79
80
81
82
83
84
85
86
87
88
89
90
91
92
93
94
95
96
97
98
99
100
101
102
103
104
105
106
107
108
109
110
111
112
113
114
115
116
117
118
119
120
121
122
123
124
125
126
127
128
129
130
131
132
133
134
135
136
137
138
139
140
141
142
143
144
145
146
147
148
149
150
151
152
153
154
155
156
157
158
159
160
161
162
163
164
165
166
167
168
169
170
171
172
173
174
175
176
177
178
179
180
181
182
183
184
185
186
187
188
189
190
191
192
193
194
195
196
197
198
199
200
201
202
203
204
205
206
207
208
209
210
211
212
213
214
215
216
217
218
219
220
221
222
223
224
225
226
227
228
229
230
231
232
233
234
235
236
237
238
239
240
241
242
243
244
245
246
247
248
249
250
251
252
253
254
255
256
257
258
259
260
261
262
263
264
265
266
267
268
269
270
271
272
273
274
275
276
277
278
279
280
281
282
283
284
285
286
287
288
289
290
291
292
293
294
295
296
297
298
299
300
301
302
303
304
305
306
307
308
309
310
311
312
313
314
315
316
317
318
319
320
321
322
323
324
325
326
327
328
329
330
331
332
333
334
335
336
337
338
339
340
341
342
343
344
345
346
347
348
349
350
351
352
353
354
355
356
357
358
359
360
361
362
363
364
365
366
367
368
369
370
371
372
373
374
375
376
377
378
379
380
381
382
383
384
385
386
387
388
389
390
391
392
393
394
395
396
397
398
399
400
401
402
403
404
405
406
407
408
409
410
411
412
413
414
415
416
417
418
419
420
421
422
423
424
425
426
427
428
429
430
431
432
433
434
435
436
437
438
439
440
441
442
443
444
445
446
447
448
449
450
451
452
453
454
455
456
457
458
459
460
461
462
463
464
465
466
467
468
469
470
471
472
473
474
475
476
477
478
479
480
481
482
483
484
485
486
487
488
489
490
491
492
493
494
495
496
497
498
499
500
501
502
503
504
505
506
507
508
509
510
511
512
513
514
515
516
517
518
519
520
521
522
523
524
525
526
527
528
529
530
531
532
533
534

Advertisement

4 Patient Alarms and INOPs
INOP Message, Indication
CCO No Press
at Information Center
CCO Not Supported
Numeric is replaced by -?-
INOP tone
CCO Overrange
CCI Overrange
Numeric is replaced by -?-
INOP tone
CCO Press Invalid
at Information Center
CCO <Press Label> Invalid
Numeric is replaced by -?-
INOP tone may sound
CCO PressOverrange
Numeric is replaced by -?-
INOP tone
CCO PulseOverrange
Numeric is replaced by -?-
INOP tone
CCO Recalibrate
Numeric is replaced by -?-
CCO/Tbl NoTransduc
Numeric is replaced by -?-
INOP tone
Tblood Overrange
Numeric is replaced by -?-
TbloodNoTransducer
Numeric is replaced by -?-
INOP tone
tcGas INOPs
INOP Message, Indication
<tcGas Label> Cal Failed
Numeric is replaced by -?-
INOP tone
<tcGas Label> Cal
Running
Numeric displays first -?-, then
numeric is displayed with a ?
<tcGas Label>
CalRequired
Numeric is replaced by -?-
INOP tone
96
What to do
CCO/CCI cannot be calculated. Make sure that the pressure chosen in the Setup CCO menu under
CCO from matches the pressure measured with the arterial catheter for CCO measurement. A
pressure from an external device cannot be used. Select another pressure label, either ABP, Ao, ART,
BAP, FAP, or UAP.
A catheter for transpulmonary C.O. measurements has been unplugged and replaced with a Right
Heart C.O. catheter, or the measurement mode has been changed manually. Silencing this INOP
switches the measurement off.
The measured CCO or CCI value is not within the specified range for CCO/CCI measurement.
The arterial pressure selected for pulse contour calculation for CCO is available but currently invalid.
Make sure the pressure transducer is connected and the zero calibration is valid.
The arterial pressure selected for pulse contour calculation for CCO is available but currently invalid.
Make sure the pressure transducer is connected and the zero calibration is valid.
The mean value of the arterial pressure values used for pulse contour calculation for CCO is below
0 mmHg or above 300 mmHg.
The pulse rate of the pressure used for pulse contour calculation for CCO is below 30 bpm or above
240 bpm.
The most recent CCO or CCI calibration was made over 8 hours ago or the arterial pressure
measurement used for CCO calculation has been zeroed after the CCO calibration was performed.
You should recalibrate CCO or CCI with transpulmonary C.O. measurements at least every 8 hours or
when the hemodynamic condition of the patient has changed. The pressure measurement must be
zeroed before a CCO calibration.
No transducer attached to the module or catheter disconnected.
Tblood out of range 17°C - 43°C.
No transducer attached to the module or catheter disconnected.
What to do
A calibration failed. Check the calibration unit, gas pressure, and tubing connections, then restart the
calibration. If the calibration has failed more than once, remembrane the transducer and restart the
calibration. If this INOP persists, contact your service personnel.
Wait until the tcpO
/tcpCO
2
2
Calibration is required before applying the transducer to the patient.
Insert a membraned transducer into the calibration chamber on the module, connect the calibration
unit to the calibration chamber, open the gas valve and start the calibration. If this INOP occurs
during a calibration, there may be a module or transducer malfunction: contact your service personnel.
calibration is finished.

Advertisement

Table of Contents

This manual is also suitable for:

Mx700Mx800Mx450Mx500Mx550Mx600

Comments to this Manuals

Symbols: 0
Latest comments: