Defibrillation
If you need to disarm the defibrillator, press the
disarms automatically when the Shock button has not been pressed within the time period specified in
the Time to Auto Disarm configuration setting.

Step 3 - Shock

Confirm that a shock is still indicated and the defibrillator is charged to the selected energy level. Make
sure no one is touching the patient or anything connected to the patient. Call out loudly and clearly,
"Stay Clear!"
If using:
WARNINGS: Defibrillation current can cause operator or bystander injury. Do not touch the patient, or equipment
connected to the patient, during defibrillation.
Alarm audio is turned off when an energy setting is selected for defibrillation, and the
message is displayed. Audio remains off until turned back on by pressing the Alarm button, Sync Mode is
turned on or the Therapy knob is turned to
Pads or switchless internal paddles - Press the flashing Shock
the HeartStart XL+ to deliver a shock.
External paddles - Simultaneously press the flashing Shock buttons located on the paddles
to deliver a shock.
Switched internal paddles - Press the orange Shock button located on the paddle to deliver
a shock.
[Cancel Charge]
soft key. Also, the defibrillator
Monitor
Pacer
or
.
6: Manual Defibrillation
button on the front of
Alarm Audio Off