Configurable Resuscitation Protocols

To monitor optional SpO
1
Once in AED Mode, attach an SpO
If you have the device configured to monitor SpO
2
pulsatile reading is obtained. For more information on SpO
page 99.
3
To activate SpO
HeartStart XL+.
Configurable Resuscitation Protocols
In AED Mode, you have the flexibility to configure the HeartStart XL+ to match your institution's
resuscitation protocols. You can:
Customize the device for the number of shocks (1-4) in a series.
Select the energy setting within a given shock series.
Set the CPR Pause interval from 1-3 minutes.
For more information, see Tables

AED Alarms

The SMART Analysis algorithm generates AED Defibrillation alarms for the conditions shown in
17. There are both audio and visual alerts, when turned on.
When monitoring a patient, the ST/AR ECG Monitoring Algorithm generates ECG alarms in AED
Mode, if turned on. See
For more information on alarms, see
Table 17
Alarm Message
Pads Off
Shock Aborted
Abnormal Shock
Dose Delivered
Pads/Paddle Type
Unknown
Equipment
Disabled: Therapy
and pulse in AED Mode and activate alarms:
2
and Pulse alarms in AED Mode, press the Alarms button on the front of the
2
34
and
"Other Alarms in AED Mode"
"Alarms"
AED Defibrillation Alarms
Condition
With pads in use, the connection
between the device and patient
has been lost.
A shock has been automatically
aborted.
Abnormal shock dose delivered
due to marginal patient
impedance.
The device detected a change in
paddles or pads type or the
therapy cable identification is
invalid.
Therapy is disabled due to an
equipment failure.
sensor to the patient (see
"Applying the Sensor"
2
in AED Mode, SpO
2
, see
2
35
in the Configuration chapter.
below.
on page 33.
Type of Alarm
High Priority
Non-Latching
Alarm
High Priority
Latching Alarm
5: AED Mode
on page 101).
monitoring begins once a
2
Chapter 9 "Pulse Oximetry"
Indication
Red Alarm message with
alarm tone
on
Table
71