Using AED Mode to Defibrillate
Table 15
AED Mode Shock Related Messages
These messages may appear when a shock is advised or immediately after a shock is delivered in AED Mode.
Audio Message
"Deliver shock now"
"Press the flashing orange
button now"
"Shock cancelled"
"Shock cancelled, No shock
advised"
"Shock cancelled. Pads
must not be touching
clothing or each other."
"Shock cancelled. Press
pads firmly to patient's bare
skin."
"Shock delivered"
"Press pads firmly to
patient's bare skin."
None
None
Display Text
Deliver Shock Now
Press Orange Button
None
Attend to Patient, Shock
Delivered
Stay Clear of Patient,
Shock Delivered
Condition
The defibrillator is charged
and waiting for you to press
the Shock button.
After charging, the device
detected a non shockable
rhythm and automatically
disarmed.
A shock was aborted due to
low impedance.
A shock was aborted due to
high impedance.
A shock has been delivered
to the patient.
Shock delivered abnormal
energy due to marginal
impedance.
The final shock in a series
has been delivered.
A shock in a
yet-to-be-completed series
has been delivered.
5: AED Mode
User Action
Press the flashing orange
Shock button.
Attend to the patient.
Confirm pad placement is
correct and pressed firmly
against the chest then try
shocking again.
Attend to the patient.
Confirm pad placement is
correct and pressed firmly
against the chest.
Attend to the patient.
Stay clear of patient.