Heart Rate and Arrhythmia Alarms
Table 8
Alarm Message
Lead/
Pads/
Paddles Off
Cannot Analyze
ECG
ECG Equipment
Malfunction
Pads ECG
Equipment
Malfunction
Equipment
Disabled: Therapy
Figure 44
For Monitor, Manual Defibrillation, Synchronized Cardioversion and Pacing
HR/Arrhythmia Technical Alarms
Condition
The multifunction electrode
pad/paddles or leads used as the
source for Wave Sector 1 during
Synchronized Cardioversion may
be disconnected or not attached
securely.
ECG data in Wave Sector 1
cannot be analyzed – an
electrode used is
disconnected/not attached
securely.
The analyzing algorithm cannot
analyze the ECG signal.
A malfunction has occurred in
the ECG hardware.
A malfunction has occurred in
the Pads ECG hardware.
Therapy is disabled due to an
equipment failure.
Basic Mode Arrhythmia Alarm Priority Chain
RED ALARMS (High Priority)
Extreme Tachy
YELLOW ARRHYTHMIA ALARMS (Medium Priority)
PVC Alarms
Beat Detection Alarms
Pacer Not Capture
Pacer Not Pacing
PVCs > x/Min.
Type of Alarm
High Priority
Non-Latching
Alarm
Asystole
V-Fib/VTach
VTach
Extreme Brady
Rate Alarms
High/Low HR
4: ECG Monitoring
Indication
Red Alarm message with
alarm tone