Continued Use

Continued Use
Once in a clinical mode, the Continued Use feature is activated. This feature facilitates continued
treatment of the same patient by retaining the current settings and patient record when the
HeartStart XL+ is turned off for less than 10 seconds. This could occur when turning the knob from
Monitor Mode to AED Mode or when the Therapy knob is accidentally moved to the
you turn the HeartStart XL+ back on within 10 seconds, it retains the most recent:
Alarm settings and conditions
Wave sector settings
Event timing
Volume settings
Vital Signs Trending data
Pacing settings
Synchronized Cardioversion settings
SpO
2
NBP value and measurement frequency
Event Summary
NOTES: The Sync feature remains active if the HeartStart XL+ is turned off for less than 10 seconds. However,
Sync is disabled when AED Mode is activated and must be turned back on upon returning to Manual
Defibrillation Mode.
The Continued Use feature does not function if both battery and external AC power are removed from
the device, even briefly.

Passwords

The HeartStart XL+ requires passwords to enter Service Mode and to make certain changes in
Configuration Mode. See the HeartStart XL+ Service Manual for the Service Mode password. The
Configuration Mode password is printed on the front of your HeartStart XL+ User Documentation
CD-ROM.