The Display

The Display
The HeartStart XL+'s display layout is easily configurable. There are four basic segments of the display.
See
Figure
Figure 27
Status
area
Parameter
area
Message area
Waveform
Display and
Soft Keys
area

Status Area

The Status Area contains key device and patient information, see
Figure 28
Date and
Time
Battery
Status
Patient information includes:
Name – The patient's name for the current event (if entered).
Date and Time – The current date and time
Paced Status – Internally paced or non-paced. The area remains blank if paced status has not
been selected. (This information is not displayed in Pacing Mode.)
Event Timer – The elapsed time for the current event, displayed in hours:minutes:seconds
Patient Category and Weight Range – Adult (defined as ≥ 25kg) or Infant/Child (defined as <
25kg). (The neonatal patient category is not supported for AED Mode and all monitoring
functions.)
27.
General Display Layout
25 Jan 2011 10:25 AM
Alarms Off
bpm
SpO2
HR
70
95
150
50
II
Pleth
Status Area
25 Jan 2011 10:25 AM
Alarms Off
Clinical Alarm Area
3: Working with the HeartStart XL+
02:27
Non-Paced
Jones, Tony
Printer Door Open
%
Pulse
bpm
NBP
q5
70
110/70
100
120
90
50
Figure
Event Timer
Paced Status
02:27
Non-Paced
Jones, Tony
Printer Door Open
Technical Alarm Area
Adult
>= 25 Kg
mmHg
10:20am
MAP
(80)
110
50
Sinus Rhythm
Dotted box
indicates
location of
menu when
displayed.
See
"Menus"
on page 31.
Menu
28.
Patient Name
Patient
Adult
Category
Weight
>= 25 Kg
Range
Multiple Technical
Alarms Indicator
27