3: Working with the HeartStart XL+

Ready For Use Indicator

The Ready For Use (RFU) Indicator is located on the right side base of the device handle. It indicates the
status of the therapy delivery functions using the following conventions:
Blinking Hourglass
Blinking red "X" with
periodic audio chirp
Blinking red "X"
without periodic
audio chirp
Solid red "X" with
periodic audio chirp
Solid red "X" without
periodic audio chirp

Power

The HeartStart XL+ is powered by a Lithium Ion battery or AC power. The battery should always be
installed so the device is ready for use whether or not AC power is available at the point of care. When
pacing, AC power should be connected if possible to prevent the battery from eventually becoming
depleted and interrupting the pacing operation.
Keep your battery charged at all times. For more information on your battery see
page 27, and
NOTES: If AC power is used as the only power source during defibrillation (for instance when no battery is
installed or when the battery is fully discharged), the HeartStart XL+ may take longer to charge to the
desired energy level, and, in the event of power loss, all settings reset to configured settings and a new
event is created when power is returned. However, all saved data remains intact (up to the device's
memory capacity) and can be found by retrieving the previous event. Keep your battery installed and
charged.
To ensure proper grounding reliability, the HeartStart XL+ must be plugged into a hospital-grade outlet.
To remove AC power, disconnect the power cord from the outlet.
If you question the AC power cord's functionality, disconnect it from the device and operate on battery
power. Replace the cord before reconnecting to AC power.
For information on power-related alarms see
24
Indicates the shock, pacing and ECG functions of the device
are ready for use. Sufficient battery power is available for
device operation.
Indicates either:
Indicates a low battery power condition but the battery is
currently charging. The device can be used but its
battery-only operation time is limited.
Indicates a critical failure has been detected that may prevent
the delivery of defibrillation therapy, pacing or ECG
acquisition.
Indicates no power available or the device cannot power on.
If, after power is returned, the indicator reverts to the
blinking black hourglass symbol, the device is ready for use.
"Battery Maintenance"
on page 166.
A low battery condition exists and the battery is not
charging.
There is no battery installed and the device is running
on AC power only.
"Power-Related Alarms"
on page 169.
Ready For Use Indicator