2: Device Basics
Removing the Battery
To remove the HeartStart XL+ Lithium Ion battery:
Push the Battery Latch to the left to eject the battery.
1
Pull on the Battery Tab and battery to completely remove the battery.
2
Figure 16
Battery Fuel Gauge
When you want to check the power remaining in your Lithium Ion battery when it is not installed in the
HeartStart XL+, press the Battery Power Gauge (see
the battery tab. Each solid green light indicates approximately 20 percent charge. A flashing green light
closest to the button indicates the battery is too weak and must be recharged before use.
Figure 17
To check the battery power remaining when the battery is inserted in the device, look at the battery
gauge on the display (see
WARNING: Use only approved batteries to power the HeartStart XL+. Use of non-approved batteries could affect
performance and results.
14
Removing the Battery
Battery Gauge
Press here
Figure 29 "Battery Charge Level"
Figure
17) located on the end of the battery opposite
on page 28).