Specifications
Water/Solids Ingress Resistance: Meets Ingress Protection level IP21.
EMC: Complies with the requirements of standard EN 60601-1-2:2002.
Safety: Meets UL 60601-1 (1st edition), EN 60601-2-4:2003, EN 60601-1:1990.
Other considerations:
The HeartStart XL+ is not suitable for use in the presence of concentrated oxygen or a flammable
anesthetic mixture with air, oxygen or nitrous oxide.
Hazards arising from software errors were minimized by the product's compliance with the software
requirements contained in EN 60601-1-4:1996.
Mode of Operation: Continuous
AC Line Powered: 100-240 VAC, 50 or 60 Hz, 1 - 0.46 A, Class I Equipment
Battery Powered: Minimum 14.4 V, Rechargeable Lithium Ion

USB Device

Correct Drive: Use the USB Drive that accompanied your device. If it is not available, use a USB device
that is USB 2.0 compatible and is ≤ 32 gigabytes.