18: Specifications
Auto Mode Repetition Time: 1, 2.5, 5, 10, 15, 30, 60 or 120 min
Maximum Measurement Time: 120 sec
Interconnect Tube Length:
M1598B Connect tubing 1.5 m (4.92 ft.)
M1599B Connect tubing 3.0 m (9.24 ft.)
Recommended Frequency of Pressure Transducer Calibration: Yearly

Patient Data Storage

Internal Event Summary: The HeartStart XL+ can store up to 8 hours of 2 continuous ECG waves, 1
pleth wave, research waves (AED Mode only) events and trending data per Event Summary. There is a
maximum capacity of approximately 50 Event Summaries of approximately 30 minutes in length.

Environmental

Temperature: 0°C to 45°C (32°F to 113°F) operating; -20°C to 70°C (-4°F to 158°F) storage
Charging the battery at temperatures above 45°C (113°F) may degrade battery life
Storing the battery for extended periods at temperatures above 40°C (104°F) reduces battery
capacity and degrades battery life.
Humidity: Up to 95% relative humidity
Printer paper may jam if the paper is wet.
Thermal printer may be damaged if wet paper is allowed to dry while in contact with printer
elements.
Atmospheric Pressure Range: Operating and Storage - 1014 mbar to 572 mbar (0 to 15,000 ft; 0 to
4,500 m).
Shock:
Operating: Half-sine waveform, duration ≤11ms, acceleration ≥ 15.3 G, 3 shocks per face.
Non-operating: Trapezoidal waveform, acceleration 30G, velocity change 7.42 m/s ±10% 1 shock per
face.
Vibration:
Frequency
(Hz)
10-100
100-200
200-2000
204
Operating Random
Slope
PSD
2
2
(dB/octave)
(m/s
)
/Hz
1.0
-3.0
0.5
Non-Operating Swept
Sine
Frequency
Amplitude Frequency
(Hz)
10-57
± .15 mm
5-500
57-150
2 g
Specifications
Non-Operating Random
PSD
Acceleration
(Hz)
0.0117
~ 2.41 Grms
2
g
/Hz