16: Troubleshooting
Table 45
ECG Monitoring Problems (Continued)
Symptom
Press "I,II.." Button to Select
Another ECG Lead
message on
display.
Lead Wire
Off
message on the
display.
Check Limb Leads
message on
the display.
Table 46
Defibrillation and Pacing Problems
Symptom
Cannot Analyze ECG
technical
alarm.
Equipment Disabled:
Therapy
technical alarm.
Connect Therapy Cable
message is displayed.
Insert Connector, Apply
Pads
message is displayed.
Check ECG Lead
message on
display during Synchronized
Cardioversion.
Energy Limited to 50J
message is displayed.
180
Possible Cause
The waveform in Wave Sector
1 is no longer valid and
another ECG source is
available.
The specified monitoring
electrode is off or not making
proper contact with the
patient.
Two or more limb lead
electrodes are off or not
making proper contact with
the patient.
Possible Cause
ECG data cannot be analyzed.
An electrode may be
disconnected or the analyzing
algorithm cannot analyze the
ECG signal.
A device failure has been
detected at startup or during
RFU testing. The device is
unable to deliver therapy.
Pads/Paddles are not connected
when attempting to charge the
device.
The Therapy cable is not
connected to the device.
Pads are not applied to the
patient.
Pads connector is not securely
connected to the Therapy
cable.
There is a leads off condition
for the synchronizing lead.
An attempt to deliver greater
than 50J using internal
paddles.
Possible Solution
Check that the monitoring electrodes/pads are
properly applied.
Use the Lead Select button to select another lead to
monitor.
Check that the monitoring electrodes are properly
applied. If necessary, prepare the patient's skin and
apply new electrodes.
Check that the limb lead electrodes are properly
applied.
Possible Solution
Check ECG signal quality. If necessary, improve lead
position or reduce patient movement.
Take the device out of use and call for service.
Connect the pads/paddles to the device.
Connect the Therapy cable to the device.
Properly apply pads to patient.
Check that pads connector and Therapy cable are
securely connected.
Check the synchronizing lead and improve the
connection.
Select a lower energy. Only energies up to 50J are
allowed using internal paddles.