Symptoms
Table 45
ECG Monitoring Problems (Continued)
Symptom
The Lead Select button does not
respond as expected.
Solid flat line - no waveform, no
Cannot Analyze ECG
leads,
alarm
Pads/Paddles Off
technical
alarm
Cannot Analyze ECG
technical
alarm.
ECG Equipment Malfunction
technical alarm.
Pads ECG Equipment
Malfunction
technical alarm.
Equipment Disabled:
Therapy
technical alarm.
Possible Cause
Device is in AED Mode.
Pads/paddles cannot be used
for the primary ECG in
Demand Mode pacing.
If a 3-Lead cable is in use or
some wires in a 5-Lead cable
are disconnected, augmented
and V-leads may not be
selectable.
Short in the patient cable or
leads.
The multifunction electrode
pads or paddles may be
disconnected or not attached
securely.
ECG data cannot be analyzed.
An electrode may be
disconnected or the analyzing
algorithm cannot analyze the
ECG signal.
A device hardware failure was
detected.
A device hardware failure was
detected.
A device failure has been
detected. The device is unable
to deliver therapy.
Possible Solution
The Lead Select button is disabled in AED Mode. To
select a lead, exit AED Mode and enter Monitor or
Manual Defibrillation Mode.
Exit pacing or choose Fixed Mode pacing.
Confirm all leads are connected.
Run an Operational Check with the ECG cable. If the
test fails, run it without the ECG cable. If the test
passes, replace the cable. If not, remove the device
from use and call for service.
Check that the pads/paddles are properly applied. If
necessary, replace the pads.
Change the ECG in Wave Sector 1 to a lead derived
from monitoring electrodes.
Check ECG signal quality. If necessary, improve lead
position or reduce patient movement.
Disconnect the ECG cable and perform an
Operational Check. If the Leads ECG test fails,
remove the device from use and call for service. If the
Leads ECG test passes, replace the ECG cable and
perform another Operational Check.
Perform an Operational Check. If the Pads/Paddles
ECG test fails with Therapy cable, disconnect the
Therapy Cable from the device when prompted in
order for the the Pads/Paddles ECG Test to run
without the cable connected. If the Pads/Paddles ECG
Test passes without the cable connected, replace the
Therapy cable. If the test fails, remove the device from
use and call for service.
Run an Operational Check. If it fails or the problem
occurs repeatedly, take the device out of use and call
for service.
16: Troubleshooting
179