Philips HeartStart XL Plus 861290 Instructions For Use Manual Page 187

Defibrillator monitor.
Hide thumbs

Advertisement

16: Troubleshooting
Table 44
General Problems (Continued)
Symptom
Non-critical Device Failure
Detected. Service Required
message on display.
Pads/Paddles Type
Unknown
message on display.
Replace Battery
message on
display.
Low Battery
message on
display.
Shutting Down in 1 min
message on display.
Shutting Down Now
message
on display.
Battery Calibration Required
message on display.
Battery Communication
Failure
technical alarm.
The power gauge on the side of
the battery does not work when
you first receive the battery.
One or more controls do not
respond as expected (e.g. Lead
Select button does not function,
soft keys do not work).
Arching or sparking during
paddles-in-pocket test therapy
discharge (Weekly Shock Test).
176
Possible Cause
A non-critical device failure
has been detected at startup or
during RFU testing.
There is a problem with the
Therapy cable and/or the
device's Therapy port.
The battery has reached its end
of life.
The battery may not have
enough remaining charge to
provide six 200J shocks and 10
minutes of monitoring.
Very low battery and the device
is not connected to AC power.
Battery charge is depleted and
the device is not connected to
AC power.
The battery needs to be
calibrated.
The HeartStart XL+ is not able
to communicate with the
battery.
Battery was shutdown to
prolong its life during
shipping.
There is a faulty control or
connection problem.
Debris or residue, such as
electrolytic gel, on the surfaces
of the paddles and/or paddles
tray.
Possible Solution
Review error and alarm messages on screen. Call for
service.
When prompted, pick the correct cable type from the
menu and press the Menu Select button. (If you select
the wrong cable type in error, unplug and re-plug in
the cable.)
Replace the Therapy cable.
If problems persist, take the device out of use and call
for service.
Replace the battery.
Connect to AC power.
Insert a fully-charged battery.
Connect to AC power to restart the device.
Connect to AC power.
Insert a fully-charged battery.
Calibrate battery before returning to service.
Connect to AC power.
Insert a fully-charged battery.
If problems persist, call for service. If battery power is
essential to patient care, take the device out of use.
Insert the battery in a HeartStart XL+ for about 1
minute to awaken it.
Remove the device from use and call for service.
Clean paddles and paddles tray.