Automated Tests
Figure 78
Auto Test Summary Results
The Automated Test Summary reports results for the hourly, daily and weekly tests that have been
performed. (See
most recent daily tests and the last 53 weekly tests. The table below describes each result and the
corresponding RFU Indicator display. For more on the RFU Indicator see
page 24.
Table 41
Result
Pass
Fail/DX
Fail/BW
Fail/CX
Fail/D
Printing/Exporting Auto Test Summaries
You can print or export the Auto Test Summaries from Operational Check. See
page 151.
Automated Test Summary
16Jan2011 1:10am
Automated Test Summary
Date and Time
Period
16Jan2011 12:11AM Hourly
16Jan2011 01:11AM Daily
15Jan2011 01:11AM Daily
14Jan2011 01:11AM Daily
13Jan2011 01:11AM Daily
12Jan2011 01:11AM Daily
11Jan2011 01:11AM Daily
10Jan2011 12:11AM Weekly
Exit
Summary
Table
41.) The AutoTest Summary lists the result of the most recent hourly test, the six
Auto Test Summary Results
RFU Indicator
Definition
Hourglass
All tests passed.
Solid red X, chirp
Service required. A
critical device failure
has been detected.
Blinking red X
Service is not required
but the battery is low
or malfunctioning.
Solid red X, chirp
An ECG cable failure
has been detected.
Hourglass
A non-critical failure
has been detected.
Result
Date and Time
Period
Pass
Pass
Pass
Pass
Pass
Pass
Pass
Pass
None.
Turn the Therapy knob to
message indicating the problem is displayed.
Refer to
Chapter 16 "Troubleshooting"
page 173 for further action.
Charge the battery as soon as possible or
replace it with a charged battery. You can
charge the battery in the HeartStart XL+ by
connecting the device to AC power.
Replace the ECG cable and rerun
Operational Check.
Turn the Therapy knob to
message indicating the problem is displayed.
Refer to
Chapter 16 "Troubleshooting"
page 173 for further action.
15: Maintenance
Result
Menu
"Ready For Use Indicator"
Required Action
Monitor
. A
on
Monitor
. A
on
"Operational Check"
on
on
165