Operational Check

Operational Check
Operational Checks supplement automated tests by verifying therapy cables, the ECG cable, paddles,
audio, charge and shock buttons, and the ability to deliver defibrillation and pacing therapy. Operational
Check also notifies you if the battery needs calibration, checks the SpO
Operational Checks should be performed at regularly scheduled intervals to supplement the hourly, daily
and weekly tests the HeartStart XL+ performs automatically.
From Operational Check you can also print Automated Test and Operational Check results.
WARNING: When performing an Operational Check or Weekly Shock Test, be sure to disconnect the ECG leads set
from the cable and confirm the HeartStart XL+ is not connected to a patient.
NOTES: Do not run Operational Check with internal paddles attached. Perform a Weekly Shock Test to test
internal paddles.
To confirm the ECG cable used during an event is functioning properly, use that same cable during
Operational Check.
If the HeartStart XL+ is equipped with multifunction electrode pads only and does not have a paddle
tray, you cannot test paddles during an Operational Check. To test paddles, you must have a simulator to
deliver the shock into. Run the Weekly Shock Test, delivering the shock into the simulator.
To enter Operational Check:
Turn the Therapy knob to
1
Press the Menu Select button.
2
3
Using the Navigation buttons, select
4
Select
menu options.
Figure 71
Confirm your exit from a clinical mode. Select Yes and press the Menu Select button.
5
The following tests are performed during an Operational Check (see
Results"
on page 155):
• General System
• Therapy Knob
• Charge button
• Leads ECG rerun (if necessary)
Monitor
Operational Check
and press the Menu Select button. See
Operational Check
Operational Check
Run Op Check
Op Check Summary
Auto Test Summary
Exit
• Sync button
• Therapy
• Leads ECG
14: Operational and Shift Checks
.
Other
and press the Menu Select button.
Run Op Check
Select
Op Check Summary
Select
Operational Checks (see
Summaries"
on page 160).
Auto Test Summary
Select
Auto Tests (see
"Auto Test Summaries"
Exit
Select
to exit the menu.
• SpO
2
• NBP
• Shock button
• Pads/Paddles ECG
and NBP modules and printer.
2
Figure 71
for Operational Check
to run an Operational Check.
to review a summary of
"Operational Check
to review a summary of
on page 161).
"Operational Check Tests and
• Printer
• Battery
• Audio
151